A Study on Improving the Consistency of the Results of Ca125 in Different Immune Detection Systems by Serum Assignment
Journal of Social Sciences and Humanities
Vol.1 , No. 1, Publication Date: Apr. 10, 2018, Page: 55-62
523 Views Since April 10, 2018, 337 Downloads Since Apr. 10, 2018
 
 
Authors
 
[1]    

Gangyi Liu, Department of Obstetrics and Gynecology in Liuzhou Municipal People’s Hospital Affiliated to Guangxi University of Science and Technology, Liuzhou, China.

[2]    

Meizhen Xie, Department of Obstetrics and Gynecology in Liuzhou Municipal People’s Hospital Affiliated to Guangxi University of Science and Technology, Liuzhou, China.

[3]    

Guosheng Su, The first affiliated hospital of jinan university;Guigang Traditional Chinese and Western Medicine Department of Orthopedics Hospital, Guigang, China.

[4]    

Lihua Qin, Guigang Traditional Chinese and Western Medicine Department of Orthopedics Hospital, Guigang, China.

[5]    

Qi Rong, Department of Obstetrics and Gynecology in Liuzhou Municipal People’s Hospital Affiliated to Guangxi University of Science and Technology, Liuzhou, China.

 
Abstract
 

Objective: The study aims to understand the Ca125 in different immune detection system of mutual recognition of the results. And it is hoped to investigate the feasibility of improving the consistency of Ca125 results in different immune detection systems by serum calibration to provide experimental basis for the mutual recognition of the results of Ca125. Methods: The Roche electrochemiluminescence assay of Ca125 mixed serum samples assignment was used as calibrator chemiluminescence immunoassay analyzer Kemei chemical plate. It is composed to be the self immune detection system with Kemei reagent and instrument. The consistency between the two detection systems before and after the assignment and the Roche electrochemiluminescence immunoassay system was compared with the reference EP9-A2. Results: After the self calibration assignment detection system and detection system of the reference correlation coefficient of r²=0.968, suggesting the good correlation between the two systems. The regression equation of Y2-X group was Y=1.029X-2.007, and relative bias E alpha =2.86% (allowable deviation EA < 25%), which showed the system is consistent with the results of the reference detection system. display after the assignment detection system with results consistent with the reference detection system. Conclusion: The difference between the results of Ca125 detection on the original detection system and the results of the reference detection system can be improved by the calibration of the serum value, as well as the consistency of Ca125 results in different immune detection systems.


Keywords
 

Serum Immune System Assignment, Detection System, Comparison, Ca125, Results of Recognition


Reference
 
[01]    

XiaoRui Luan, Weiping Li and Wenshen Di. Progress of ovarian cancer research of Serum tumor markers for early diagnosis [J]. International Journal of Obstetrics and Gynecology. 2009, 36 (6): 458-461.

[02]    

Qiaoti Zhang, Erfu Xie and Yan Zhang, et al. Consistency evaluation of three CA125 detection systems [J]. Journal of Clinical Pathology, 2015, 35 (1): 65-69.

[03]    

CHEN Qubo, HUANG Wujiao, LI Cuicui, et al. Comparison research of Array-ELISA and ECLIA in determination of six tumor markers in serum [J]. Laboratory Medicine, 2012, 27 (1): 4-7.

[04]    

National Committee for Clinical Laboratory Standards. Medthod Comparison and Bias Estimation Using Patients Samples, Approved Guideline EP09-A2 [S]. Ann Arbor: Thomson Reuters, 2002.

[05]    

Honghua Sun, Lingbo Shi and Hong Kang, et al. NCCL SEP10-A2 files were used to evaluate the performance of self built and biochemical testing system [J]. Laboratory Medicine 2010, 25 (5): 345-348.

[06]    

Jie Chen, Lanlan Wang and Lixin Wang, et al. Evaluation of the consistency of the 2 methods based on NCCLS-EP9-A2 for the analysis of Luminescent Immunoassay [J]. Journal of Clinical Laboratory, 2006, 24 (3): 169-171.

[07]    

Honghua Sun, Lingbo Shi and HongKang, et al. A study on the value of fresh serum transfer to improve the results of self built and biochemical testing system two [J]. International Journal of Laboratory Medicine, 2011, 32, (2), 156-158.

[08]    

WANG Li,QUAN Cui-xia,NIU Lu-lu,et al. Improvement of testing results from various biochemical analytical systems by transfer of value assignment with fresh sera [J]. Chinese Journal Of Clinical Laboratory Science. 2008, 26 (4): 294-296.

[09]    

Wuying Ye, Liusan Wang and Ziyu Shen. National clinical laboratory operating procedures [M]. Beijing: People's Medical Publishing House, 2015, 539-540.

[10]    

Lier Wang and Chengju Ou. Comparative study on the results of the self determination of immune detection system [J]. International Journal of Laboratory Medicine, 2011, 32 (9): 957-960.

[11]    

Pritzker KH. Cancer biomarkers: easier said than done. Clin Chem, 2002, 48: 1147-1150.

[12]    

Fleisher M, Dnistrian Am, Sturgeon CM, et al. Practice guidelines and recommendations for use tumor markers in the clinic. Washington DC: AACC Press, 2002. 20-25.

[13]    

Xiong Zou. Effects of tumor markers on early diagnosis of cancer [J]. Chinese Journal Of Laboratory Medicine. 2002, 25 (2): 71-72.

[14]    

TIAN Man-fu HAN Bo. The clinical value of the detections of CA199, CA125, CA153 and CEA in malignant tumors [J]. Journal Of Clinical And Experimental Medicine. 2010, 9 (7): 483-485.

[15]    

Lloyd KO, Yin BW, Kudryashov V. Isolation and characterization of ovarian cancer antigen CA 125 using a new monoclonal antibody (VK-8): identification as a mucin-type molecule [J]. Int J Cancer, 1997, 71 (5): 842-850.

[16]    

Yuhua Ji, Yixun Tao. The operationg factors affecting the results of enzyme linked immunosorbent assay [J]. Laboratory Medicine. 2006, 21 (z1): 47-49.

[17]    

Francesco Dati Jill Tate. Traceability, international standardization programs, and world-wide reference systems in laboratory medicine. Chinese Journal Of Laboratory Medicine. 2008, 31 (3): 250-252.

[18]    

LI Jin-ming. Standardization of molecular diagnosis in clinical laboratory [J]. Chinese Journal Of Laboratory Medicine. 2006, 29 (6): 483-486.

[19]    

Jialin Wu, Hong shun Tang. Comparison of enzyme linked immunosorbent assay with conventional method for quantitative detection of tumor marker CEA [J]. Experimental And Laboratory Medicine. 2010, 28 (2): 143-144.





 
  Join Us
 
  Join as Reviewer
 
  Join Editorial Board
 
share:
 
 
Submission
 
 
Membership